1. Skip to Menu
  2. Skip to Content
  3. Skip to Footer>

Color Mono II Test Procedure

 Mono_Procedure_Interpretation_1 STEP 1: Using a fresh stirrer pipet each time, add one drop (0.05ml) of each sample onto a separate circle on the test card. Repeat by adding one drop of REACTIVE or NONREACTIVE CONTROL from the dropper vials supplied.
 Mono_Procedure_Interpretation_2 STEP 2: Shake the vial of Color Mono Latex Reagent gently to uniformly mix the suspension. Add one free-falling drop of reagent to each control and sample.
 Mono_Procedure_Interpretation_3 STEP 3: Using the flat end of the stirrer pipets, mix each sample and reagent and spread over the entire cirlce.
 Mono_Procedure_Interpretation_4 STEP 4: Gently rotate the card for 2 minutes. Observe for agglutination under high intensity light. All test results should be compared to both REACTIVE and NONREACTIVE CONTROLS.
     

Interpretation

REACTIVE

Any degree of agglutination or rimming within the test area as compared to the nonreactive control.

NONREACTIVE

Smooth or finely granular suspension with no visible agglutination.
 Mono_Interpretation

MONO Procedure

Test Kit Size

Catalog Number

25 Tests

450025

50 Tests

450050

100 Tests

450100

1000 Tests

4501000

CPT Code  86308

510(k)-K821386