Color Mono II Test Procedure |
|||
![]() |
STEP 1: | Using a fresh stirrer pipet each time, add one drop (0.05ml) of each sample onto a separate circle on the test card. Repeat by adding one drop of REACTIVE or NONREACTIVE CONTROL from the dropper vials supplied. | |
![]() |
STEP 2: | Shake the vial of Color Mono Latex Reagent gently to uniformly mix the suspension. Add one free-falling drop of reagent to each control and sample. | |
![]() |
STEP 3: | Using the flat end of the stirrer pipets, mix each sample and reagent and spread over the entire cirlce. | |
![]() |
STEP 4: | Gently rotate the card for 2 minutes. Observe for agglutination under high intensity light. All test results should be compared to both REACTIVE and NONREACTIVE CONTROLS. | |
Interpretation |
|||
REACTIVE |
Any degree of agglutination or rimming within the test area as compared to the nonreactive control. | ||
NONREACTIVE |
Smooth or finely granular suspension with no visible agglutination. | ||
![]() |
|||
MONO Procedure
|
Test Kit Size |
Catalog Number |
|
25 Tests |
450025 |
|
50 Tests |
450050 |
|
100 Tests |
450100 |
|
1000 Tests |
4501000 |
CPT Code 86308