ASI RPR CARD TEST PROCEDURE

 RPR_Procedure_Interpretation_1
STEP 1: Using a fresh pipet, add one drop of serum or plasma onto a circle on the test card. Repeat by adding a drop of REACTIVE, WEAK REACTIVE, or NONREACTIVE CONTROL from the dropper vials.
 RPR_Procedure_Interpretation_2
STEP 2: Using the flat end of the stirrer pipet, spread the sample over the entire area of the test circle.
 RPR_Procedure_Interpretation_3
STEP 3: Attach the needle to the dropping bottle. Mix the Carbon Antigen. Dispense several drops into the dropping bottle cap to ensure a clear needle passage.
 RPR_Procedure_Interpretation_4 STEP 4: Dispense one drop of the antigen suspension onto each sample. DO NOT RESTIR the sample and the antigen. Aspirate any antigen from the bottle cap.
 RPR_Procedure_Interpretation_5
STEP 5: Place the card on an automatic rotator and cover for 8 minutes. When complete, physically rotate/tilt the test card 3-4 times.
 RPR_Procedure_Interpretation_6 STEP 6: Immediately read results macroscopically in the “wet” state under a high intensity light source.

INTERPRETATION

 RPR_Procedure_Interpretation_Results

RPR Procedure

Test Kit Size

Catalog Number

25 Test Kit

900025

100 Test Kit

900100

500 Test Kit

900500

500 Test Kit
(15 Well Cards)

900500-15
500 Test Kit
(15 Well Cards, W/Out Controls)
987115

5,000 Test Kit
(10 Well Cards)

9005000-10

5,000 Test Kit
(15 Well Cards)
9005000-15
5,000 Test Kit
(15 Well Cards, W/Out Controls)
987515

5,000 Test Kit
(30 Well Cards)

9005000-30

10,000 Test Kit
(10 Well Cards)

90010000-10

10,000 Test Kit
(15 Well Cards)
90010000-15
10,000 Test Kit
(15 Well Cards, W/Out Controls)
987X15

10,000 Test Kit
(30 Well Cards)

90010000-30


CPT Code 86592

510(k)-K851504